Correlation between the prothrombin time values obtained by point-of care prothrombin time meter and clinical laboratory in paediatric intensive care unit of a developing country
DOI:
https://doi.org/10.18203/2349-3291.ijcp20242330Keywords:
Laboratory, PICU, Plasma, POC, PTAbstract
Background: Monitoring of prothrombin time (PT) is essential for adequate management of children in pediatric intensive care unit (PICU) who require serial and repeated PT measurements. In contrast to laboratory testing, point-of-care (POC) PT meters have been configured especially for bedside that are cost effective with rapid turnaround time. Our study aimed to assess its feasibility for routine use in PICU by assessing the correlation between PT values obtained through the POC PT meter and those acquired from the laboratory.
Methods: A total of 200 children of either gender aged between 1 month to 12 years, admitted to PICU of Indus hospital Karachi Pakistan from February to August 2023 were analysed in this cross-sectional study. Venous blood samples were collected in PT bottles and transported to the pathology department within a time frame of 20-30 minutes for analysis. Concurrently, a drop of blood was obtained via finger prick and promptly assessed using the POC device-Coaguchek INRange system, (Roche diagnostics Ltd, USA). Pearson’s correlation was applied.
Results: 200 children (96 males, 104 females) having a mean age of 6.42±3.55 years. The mean PT values with POC meter and Laboratory were 2.588±1.299 and 2.574±1.304 respectively (p=0.100). Very strong correlation of PT between POC PT meter and laboratory evaluation was observed (r=0.996, p<0.001).
Conclusions: The strong correlation observed between POC PT meter and laboratory supports the use of these devices for timely coagulation monitoring, potentially improving patient outcomes in PICUs.
Metrics
References
Iba T, Helms J, Connors JM, Levy JH. The pathophysiology, diagnosis, and management of sepsis-associated disseminated intravascular coagulation. J Intensive Care. 2023;11(1):24.
Kunwar S, Alam M, Ezekwueme F, Yasir M, Lawrence JA, Shah S, et al. Diagnostic Scores and Treatment Options for Acute Disseminated Intravascular Coagulation in Children. Cureus. 2021;13(9):e17682.
Huybrechts KF, Gopalakrishnan C, Franklin JM, Zint K, France LR, Bartels DB, et al. Claims Data Studies of Direct Oral Anticoagulants Can Achieve Balance in Important Clinical Parameters Only Observable in Electronic Health Records. Clin Pharmacol Ther. 2019;105(4):979-93.
Lakshmy R, Kumar S. Comparative evaluation of Point Oof Care coagulation monitoring by coaguchek XS-comparison with standard laboratory method. Indian J Thorac Cardiovasc Surg. 2010;26:125-8.
Williams NX, Carroll B, Noyce SG, Hobbie HA, Joh DY, Rogers JG, et al. Fully printed prothrombin time sensor for point-of-care testing. Biosens Bioelectron. 2021;172:112770.
Tonthong S, Rungpupradit J. Coagulation testing: Comparison of portable (CoaguChek® XS) and automated coagulation analyzer in healthy cats. Vet World. 2020;13(11):2541-5.
Christensen TD, Larsen TB. Precision and accuracy of point-of-care testing coagulometers used for self-testing and self-management of oral anticoagulation therapy. J Thromb Haemost. 2012;10:251-60.
Greenway A, Ignjattovic V, Summerhayes S, Newall F, Burgess J, Derosa L. Point of care monitoring of oral anticoagulation therapy in children. Comparison of the CoaguChek system with venous INR and venous INR using an International Referance Thromboplastin preparation. Thromb Haemostat. 2009;159-65.
Christensen TD, Larsen TB, Jensen C, Maegaard M, Sorensen B. International normalised ratio (INR) measured on the CoaguChek S and XS compared with the laboratory for determination of precision and accuracy. Thromb Haemost. 2009;101:563-9.
Shakil U, Usman N, Waqar T, Qamar A. Correlation between prothrombin time levels obtained using prothrombin meter with laboratory in neonates. Pak Armed Forces Med J. 2019;69(5):1083-7.
Christensen TD, Larsen TB. Precision and accuracy of point-of-care testing coagulometers used for self-testing and self-management of oral anticoagulation therapy. J Thromb Haemost. 2012;10(2):251-60.
Van den Besselaar AM. Accuracy, precision, and quality control for point-of-care testing of oral anticoagulation. J Thromb Thrombolysis. 2001;12(1):35-40.
Ottawa (ON): Canadian Agency for Drugs and Technologies in Health. Point-of-Care Testing of International Normalized Ratio for Patients on Oral Anticoagulant Therapy: Systematic Review and Economic Analysis. 2014.
Kalçık M, Yesin M, Gürsoy MO, Gunduz S, Karakyyun S, Astarcioglu MA, et al. Comparison of the INR Values Measured by CoaguChek XS Coagulometer and Conventional Laboratory Methods in Patients on VKA Therapy. Clin Appl Thromb Hemost. 2017;23(2):187-94.
Expert WHO Committee on Biological Standardization. Guidelines for thromboplastins and plasma used to control oral anticoagulation therapy. World Health Organ Tech Rep Ser. 1999;889:64-93.
Lawrie AS, Hills J, Longair I, Green L, Gardiner C, Marchin SJ, et al. The clinical significance of differences between point-of-care and laboratory INR methods in over-anticoagulated patients. Thromb Res. 2012;130(1):110-14.
Medical Advisory Secretariat. Point-of-Care International Normalized Ratio (INR) Monitoring Devices for Patients on Long-term Oral Anticoagulation Therapy: An Evidence-Based Analysis. Ont Health Technol Assess Ser. 2009;9(12):1-114.
Ryan F, O'Shea S, Byrne S. The reliability of point-of-care prothrombin time testing. A comparison of CoaguChek S and XS INR measurements with hospital laboratory monitoring. Int J Lab Hematol. 2010;32(1 Pt 1):e26-33.
Rajsic S, Breitkopf R, Bachler M, Treml B. Diagnostic Modalities in Critical Care: Point-of-Care Approach. Diagnostics (Basel). 2021;11(12):2202.
Watkins LA, Dial SP, Koenig SJ, Kurepa DN, Mayo PH. The Utility of Point-of-Care Ultrasound in the Pediatric Intensive Care Unit. J Intensive Care Med. 2022;37(8):1029-36.
Patel K, Suh-Lailam BB. Implementation of point-of-care testing in a pediatric healthcare setting. Crit Rev Clin Lab Sci. 2019;56(4):239-46.