Bubble continuous positive airway pressure therapy for neonatal respiratory distress at birth in level III newborn unit in Chengalpattu Medical College Hospital: a prospective observational study

Diana Grace R., Suresh Kumar S., Sathya .

Abstract


Background: Bubble continuous positive airway pressure (bCPAP), a non-invasive respiratory device provides continuous pressure that helps recruitment of more alveoli, increases the lungs, functional residual capacity and decreases the work of breathing in newborns admitted with respiratory distress. Bubble continuous positive airway pressure (bCPAP) is the most important respiratory support in different types of respiratory conditions in level III. In this observational study, author reported this research using bCPAP therapy as the primary respiratory support in level III unit in tertiary care centre in Chengalpattu, Tamil Nadu, India. Despite reporting their indications, duration of use and adverse effects we tried to search for further improvement in other areas of CPAP therapy in level III newborn unit.

Methods: This prospective observational study included 250 babies delivered in obstetric unit and admitted with respiratory distress within 6 hours of birth at level III care. They were treated according to level III newborn unit protocol. Those data were collected and entered in the proforma. Newborns were followed up till discharge.

Results: A total of 250 babies satisfying the inclusion criteria delivered in Chengalpattu Medical College Hospital, Tamil Nadu, India (mean gestational age 36±2 weeks, mean birth weight 2.5±1.2 kg were included. All newborns were given bCPAP as the primary support. The most common underlying cause of respiratory distress was transient tachypnea of newborn (44%), followed by respiratory distress syndrome (24%) and prolonged respiratory transition (18%). The therapy success rate was 86%. Only 35 newborns failed to respond to CPAP. The most common adverse effect was eye puffiness (19%).

Conclusions: Bubble continuous positive airway pressure (b CPAP) therapy use is being well established in level III unit for various respiratory conditions with minimal failure and adverse effects. Its use in extreme preterms and initiation after 6 hours is controversial.


Keywords


CPAP therapy, Neonatal unit, Respiratory distress

Full Text:

PDF

References


Millar D, Kirpalani H. Benefits of noninvasive ventilation. Ind Pediatr. 2004;41(10):1008-11.

Gregory GA, Kitterman JA, Phibbs RH, Tooley WH, Hamilton WK. Treatment of the idiopathic respiratory distress syndrome with continuous positive airway pressure. N Engl J Med. 1971;284 (24):1333-40.

Nowadzky T, Pantoja A, Britton JR. Bubble continuous positive airway pressure, a potentially better practice, reduces the use of mechanical ventilation among very low birth weight infants with respiratory distress syndrome. Pediatrics. 2009;123(6):1534-40.

Upadhyay A, Deorari AK. Continuous positive airway pressure- a gentler approach to ventilation. Ind Pediatr. 2004;41(5):459-69.

Rezzonico R, Caccamo LM, Manfredini V. Impact of the systematic introduction of low cost bubble nasal CPAP in a NICU of a developing country: a prospective pre and post intervention study. BMC Pediatr. 2015;15:26.

Dewez JE, Van den Broek N. Continuous positive airway pressure to treat respiratory distress in newborns in low and middle income countries. Trop Doct. 2017;47(1):19-22.

Jayashree M, Babu KHB, Singhi S, Nallasamy K. Use of nasal bubble CPAP in children with hypoxemic clinical pneumonia-report from a sourfce limited set up. J Trop Pediatr. 2016;62(1):69-74.

Abdulkathir I, Hassan L, Abdullahi F, Purdue S, Ogala WN. Nasal bubble CPAP: one year experience in a neonatal unit of a tertiary health facility in northwestern Nigeria. Niger post grad Med J. 2015;22(1):21-4.

Cummings JJ, Polin RA. The committee on fetus and newborn, American academy of pediatrics. noninvasive respiratory support. Pediatrics. 2016;137(1):169-73.

Bassiouny MR, Gupta A, EI Bualy M. Nasal continuous positive airway pressure in the treatment of respiratory syndrome: an experience from a developing country. J Trop pediatr. 1994;40 (6):341-4.

Gallacher DJ, Hart K, Kotecha S. Common respiratory conditions of newborn. Breathe (shef). 2016;12(1):30-42.

Dehdashtian M, Aletayeb M, Malakian A, Aramesh MR, Malvandi H. Clinical course in infants diagnosed with transient tachypnea of newborn: a clinical trial assessing the role of conservative versus conventional management. J Chin Med Assoc. 2018;81(2):183-6.

Sinha SK, Donn SM. Fetal to neonatal maladaptation. Semin Fetal neonatal Med. 2006;11(3):166-73.

Reuter S, Moser C, Baack M. Respiratory distress in the newborn. Pediatr Rev. 2014;35(10):417-28.

Roberts CT, Davis PG, Owen LS. Neonatal noninvasive respiratory support: synchronized NIPPV, non-synchronised NIPPV or bilevel CPAP: what is the evidence in 2013? Neonatol. 2013;104(3):203-9.

Kawaza K, Machen HE, Brown J. Efficacy of a low cost bubble CPAP system in treatment of respiratory distress in a neonatal ward in Malawi. PLoS One. 2014;9(1):e86327.

Al-lawama M, Abdelghani T, Badran E. Mortality and short term outcomes of very low birth weight infants at a tertiary care centre in Jordan: comparison with other countries. Pediatr Ther. 2017;7:311.

Askin DF. Fetal to neonatal transition-what is normal and what is not? Neonatal Netw. 2009;28(3):e33-e40.

Hilman NH, Kallapur SG, Jobe AH. Physiology of transition from intrauterine to extrauterine life. Clin Perinatol. 2012;39(4):769-83.

Gupta S, Donn SM. Continuous positive airway pressure: to bubble or not to bubble? Clin Perinatal. 2016;43(4):647-59.

Liu J, Wang Y, Fu W, Yang CS, Huang JJ. Diagnosis of neonatal transient tachypnea and its differentiation from respiratory distress syndrome using lung ultrasound. Med (Baltimore). 2014;93(27):197.

Osman AM, El- Farrash RA, Mohammed EH. Early rescue neopuff for infants with transient tachypnea of newborn: a randomized controlled trial. J Matern Fetal Neonatal Med. 2017;16:1-7.

Celebi MY, Alan S, Kahvecioglu D. Impact of prophylactic continuous positive airway pressure on transient tachypnea of newborn and neonatal intensive care admission in newborns delivered by elective cesarean section. Am J Perinatol. 2016;33(1):99-106.

Yagui AC, Vale LA, Haddad LB. Bubble CPAP vesus CPAP with variable flow in newborns with respiratory distress: a randomized controlled trial. J Pediatr (Rio J). 2011;87(6):499-504.

Rubarth LB, Quinn J. Respiratory development and respiratory distress syndrome. Neonatal Netw. 2015;34(4):231-8.

Mahoney AD, Jain L. Respiratory disorders in moderately preterm, late preterm, and early term infants. Clin Perinatol. 2013;40(4):665-78.

Jones M. Effect of preterm birth on airway function and lung growth. Pediatr Respir Rev. 2009;10(1):9-11.

Miller MJ, Difiore JM, Strohl KL, Martin RJ. Effects of nasal CPAP on supraglottic and total pulmonary resistance in preterm infant. J Appl Physiol. 1990;68(1):141-6.

Bamat N, Jensen EA, Kirpalani H. Duration of continuous positive airway pressure in premature infants. Semin Fetal Neonatal Med. 2016;21(3):189-195.

Hanin M, Nuthakki S, Malkar MB, Jadcherla SR. Safety and efficacy of oral feeding in infants with BPD on nasal CPAP. Dysphagia. 2015;30(2):121-7.

Imbulana DI, Manley BJ, Dawon JA, Davis PG, Owen LS. Nasal injury in preterm infants receiving non-invasive respiratory support: a systematic review. Arch Dis Child Fetal Neonatal Ed. 2018;103(1):29-35.